P R O D U C T S
PWM-60 Fizeau Interferometric Wavelength Meter
PWM‑60 adopts the Fizeau interferometry principle. It converts laser wavelength information into spatial interference fringes, and obtains wavelength readings via detector acquisition and algorithm demodulation. The product supports 8‑channel measurement and 8‑channel laser wavelength locking feedback. Targeting light source tuning, status monitoring and wavelength control in multi‑laser experiments, it provides measurement data and locking feedback.
Features
- Static optical path, no mechanically moving components
- High resolution to resolve subtle wavelength variations
- USB‑powered, plug‑and‑play operation
- 8‑channel measurement and wavelength‑locking‑loop feedback
Application
- Cold‑atom and quantum experiments
- Laser spectroscopy and atomic‑molecular research
- R&D and production testing of lasers
- Optical sensing and remote sensing
Main parameters
|
Specifications |
|
|
Model |
PWM-60 |
|
Measuring wavelength range |
350-1100nm |
|
Absolute accuracy |
350-750nm:60 MHz 750-1100nm:30MHz |
|
Measurement resolution |
≤400 kHz |
|
Long‑term drift |
<5 MHz/day |
|
Single‑channel measurement rate |
≥1000 Hz |
|
Input fiber & optical interface |
Single‑mode fiber, FC/PC |
|
Communication interface & SDK |
USB 3.0 / C++ |
|
Locking‑feedback output interface |
BNC |
|
External trigger & synchronization |
TTL |
|
Power supply & power consumption |
USB power supply / 3 W |
|
Options |
|
|
Optical switch |
8‑channel supported / with wavelength‑locking feedback function |
|
780 nm frequency‑stabilized laser |
For calibrating long‑term drift error |
Dimensions: 228 × 385.5 × 127.7 mm